Procedure - Solid Samples
Determine the instrument blank.
Repeat steps 2b through 2g a minimum of three times.
Set the blank following the procedure outlined in the operator's instruction manual.
Instrument calibration/drift correction.
Login a minimum of 3 Standard replicates.
Weigh approximately *** grams of a calibration/drift standard, enter the mass and standard identification into appropriate replicate fields.
Note: LECO Reference Materia***o not require preparation.See preparation statement on the reference material certificate.
可选具有压力补偿和温度控制性能的标气检测气路。广东美国LECO氧氮氢分析仪采购
Procedure – Powder/Chip Samples
Prepare the instrument as outlined in the operator's instruction manual.
Login a minimum of three Blank reps.
Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open. Insert a *** Nickel Capsule (leave capsule open) into the open port at top of loading head.
Note: samples using automation should be placed in the appropriate autoloader position before starting the analysis sequence. Once the sequence has started, the automatic analysis will start and end automatically.
Determine the instrument blank.
a. Login a minimum of 3 Blank replicates.
b. Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open.
c. Press the Analyze button on the instrument screen again, the loading head slide-block will close and the lower electrode will open.
d. Clean the upper and lower electrode manually, or, if applicable, remove thecrucible and press the analyze button to clean with the automatic cleaner.
e. Firmly place a graphite crucible on the lower electrode tip.
f. Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and endautomatically.
g. Repeat steps 2b through 2f a minimum of three times.
h. Set the blank following the procedure outlined in the operator's instruction manual.
氧氮氢分析仪是快速、准确分析氧、氮、氢元素的、高性能仪器,整机采用模块化、一体化设计,即主机由可靠的脉冲电极炉、高灵敏高精度的检测系统、布局合理和气密性好的气路系统、可靠耐用的电源系统、稳定的电路系统、高校节能的水循环散热系统等六个**的硬件模块单元组成。
氧氮氢分析仪的测量原理:
测量范围非常***。分析样品时,样品被称重后放到样品端口中,然后用载气进行冲洗,防止大气(含氧气和氮气)进入到炉子系统中。
石墨坩埚在脉冲炉中脱气,以尽量减少自带的污染。经过稳定阶段后样品落入坩埚中并熔融。样品中的氧与石墨坩埚中的碳反应生成一氧化碳。氮和氢以单质的形式释放。载气和样品气体通过粉尘过滤网后再进入到氧化铜催化炉中,将一氧化碳氧化成二氧化碳。
二氧化碳进入到红外池中用于测定氧含量,被测后的气体导入化学试剂管,这时二氧化碳和水被化学试剂除去,氮元素通过热导检测池含量被测定。测量氢时,载气换为氮气,样品气体通过舒茨试剂而不是氧化铜催化剂。测氧和氮时也可采用更廉价的氩气作为载气。 稳固的炉头试样加载结构。
氧氮氢分析仪的常见故障及解决方法:
氧和氮空白值超过20。这是由于气流小,不能将炉子中的空气驱赶出去。可调节气体流量,调节载气压力在0.2~0.4MPa。接通仪器载气,放一个石墨坩埚在下电极上,打开主电源开关,点击软件上的关炉按钮,关闭炉子并等待10s。调节流量调节器,直到流量计a显示为30L/h,打开炉子。调节调节器直到流量计b显示为50L/h,再次关闭炉子。如果以上设置不稳定,则增加流量至100L/h,反复调节直至仪器稳定。 采用三个**的红外器检测氧含量,同时检测CO和CO2,然后在转化为CO2集中检测。新疆元素分析仪价格
力可创新的批次和重复样品数据分组功能极大简化了样品结果输出和结果标准偏差的自动计算。广东美国LECO氧氮氢分析仪采购
ONH836氧氮氢分析仪
高频燃烧红外检测技分析无机材料中的碳、硫含量。
采用惰性气体保护熔融法,提供各种无机材料、钢铁、有色金属、硬质合金、难熔金属、陶瓷等材料中宽范围的氧氮氢同时分析。
该仪器配置MS Windows®的软件接口,专为触摸屏操作定制。长期的开发研究与客户反馈相结合,这个简单方便的软件使得用户可以快速完整访问仪器控制,分析设置,故障诊断,数据报告, **提高日常分析的效率并节省用户宝贵的时间。 可选的Cornerstone® Mobile软件功能,使得用户可以在手机,便携式电脑及台式电脑上便捷的访问仪器数据,分析曲线图形以及仪器的状态,实现对仪器的遥控。
广东美国LECO氧氮氢分析仪采购
上海禹重实业有限公司致力于仪器仪表,是一家服务型公司。公司业务涵盖光谱仪,色谱仪,质谱仪,样品前处理等,价格合理,品质有保证。公司将不断增强企业重点竞争力,努力学习行业知识,遵守行业规范,植根于仪器仪表行业的发展。禹重科技立足于全国市场,依托强大的研发实力,融合前沿的技术理念,飞快响应客户的变化需求。