Instrument calibration/drift correction.
a. Login a minimum of 3 Standard replicates.
b. Weigh a ~1.000 gram of a calibration/drift standard, enter the mass and standard identification into appropriate replicate fields.
c. Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open.
d. Place the calibration/drift standard into the open port at the top of the loading head.
e. Press the Analyze button on the instrument screen again, the loading head slide-block will close and the lower electrode will open.
可选六轴可旋转触摸屏幕,增强了操作的工效和直观性。湖南美国LECO氧氮氢分析仪优惠价格
Determine the instrument blank.
a. Login a minimum of 3 Blank replicates.
b. Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open.
c. Press the Analyze button on the instrument screen again, the loading head slide-block will close and the lower electrode will open.
d. Clean the upper and lower electrode manually, or, if applicable, remove thecrucible and press the analyze button to clean with the automatic cleaner.
e. Firmly place a graphite crucible on the lower electrode tip.
f. Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and endautomatically.
g. Repeat steps 2b through 2f a minimum of three times.
h. Set the blank following the procedure outlined in the operator's instruction manual.
北京美国LECO氧氮氢分析仪采购20位或100位全自动分析系统。
氧氮氢分析仪的使用注意事项:
1 、氧氮氢分析仪在初次启用前,应该对连接点、焊点、阀门等进行检漏,以确保空气中的氧不会反渗进入管道及仪器内部,造成测量数值偏高。
2、再次使用仪器前,要进行管道系统净化,将漏入的空气除干净,同时确保连接取样管路时没有漏入空气。
3、样气中氧含量的变化会受管道材质及表面粗糙度影响,因此一般连接管路选用铜管或抛光过的不锈钢管,而不用塑料管、橡胶管等。
4、微量元素分析时,要避免各种管件、阀门、表头等死角对样气造成污染,因此必须尽可能的简化气路系统,连接件死角要小,以防止溶解氧逸出造成污染,使用水封、油封及腊封等设备,才能较好的确保数据的。
美国力可LECO氧氮氢分析仪
公司介绍:
美国力可公司(LECO)始于1936年 ,以生产用于冶金、机械、能源、化工、农业、食品及科研开发领域的各种分析仪器。
ONH836氧气/氮气/氢气元素分析仪设计用于使用惰性气体熔融技术对无机材料,铁和非铁合金以及耐火材料中的氧,氮和氢含量进行宽范围的测量。
坚固的样品加载区域可简化日常维护并大程度减少大气污染
用于氧气检测的三重红外电池可实现业内大的检测范围
通过自动化选项提高实验室生产率
可用的自动清洁器大程度地减少了两次分析之间的手动清洁需求
适用于坩埚和样品的20位往复式装载机
安装在动臂上的触摸屏界面可提供更好的人体工程学设计和直观的操作
红外(IR)和热导率(TC)检测器,没有活动部件,也无需手动调节
改善TC电池的稳定性
可绕过的OMI-2载气洗涤塔,提高了TC电池的稳定性
获得Dyanamic流量补偿可进一步增强TC细胞的稳定性
上海禹重给您提供各种氧氮氢分析仪,欢迎来电咨询!
Method Selection
Two methods are described in this application note; either method can be used to analyze iron, steel, nickel-, and cobalt-base alloys. The Precision Method is recommended for general use and will provide the best precision and accuracy throughout the typical O, N, and H concentrations found in this group of metals; approximate cycle time is 3.5 minutes. The Fast Track method can be used where speed of analysis is a critical component; for example, when molten metal is being sampled and results are required in the shortest possible time. This method will produce suitable results for most samples; approximate cycle time is 2.25 minutes. As noted above, sampling and sample preparation are key elements to accurate O, N, and Hdetermination as well. It is up to the user to determine which method best meets their needs.
通用下电极坩埚支座,容纳各种坩埚使用。安徽氧氮氢分析仪咨询报价
OHN836系列仪器采用惰性熔融法。湖南美国LECO氧氮氢分析仪优惠价格
Procedure - Solid Samples
Instrument calibration/drift correction.
Place the calibration/drift standard in a *** Nickel Basket, and if applicable, place the sample into the appropriate autoloader position.
Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open.
Note: samples using automation should be placed in the appropriate autoloader position before starting the
analysis sequence. Once the sequence has started, the automatic analysis will start and end automatically.