氧氮氢分析仪的操作过程

氧氮氢分析仪可根据的用户的要求定制。该分析仪配备两个不同灵敏度的探测器,在更多的应用中保证了准确度。

操作简单而且安全。在样品称重以后,样品的重量数据通过电子界面传导至相连的电脑中,也可以在软件界面中手动输入样品的重量。

样品被放置在水平放置的炉子中的低温区域。分析开始以后,炉子开始向上转动,然后样品逐渐落入加热区。通过添加载体气体氮气,氢气被扩散出来并且随载体被传导至一个高灵敏度的热传导测量池。

通常一次测量的时间在3到15分钟之间。在分析的过程中设备可以显示测试信号和设备的参数,从信号的评估到结果在界面上的显示全部自动完成;而测量数据则被传导至LIMS(实验室信息管理系统)中。无需过多的保养和维护,而除尘器和净化管的维护工作也方便。


减少了繁琐的额外数据处理。山东元素分析仪咨询报价

Procedure – Powder/Chip Samples

Determine the instrument blank.


Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically.


Repeat steps 2b through 2g a minimum of three times.


Set the blank following the procedure outlined in the operator's instruction manual.


Instrument calibration/drift correction.

Login a minimum of three Standard reps.


Weigh approximately *** grams of a calibration/drift standard into a *** Nickel Capsule; enter the mass and standard identification into appropriate rep fields. If applicable, place the nickel capsule in the appropriate autoloader position.


Note: LECO Reference Materia***o not require preparation. See preparation statement on the reference material certificate. Solid Standards may be used to calibrate when chip or powder standards are not available.




山西分析仪供应商可选具有压力补偿和温度控制性能的标气检测气路。

LECO—Your source for total analytical solutions


CS CS 844 and 744 Series: Carbon/Sulfur by Combustion

Carbon and sulfur determination by combustion infrared detection

Quick, accurate, and affordable determination for both production control and research

Calibration, analysis, evaluation, and diagnostic functions accessible via user-friendly brand software



GDS GDS 500A and 850A Atomic Emission Spectrometers

Spectral range from 120 to 800 nm

Continuous profile of concentration/thickness

Large, dynamic range with concentrations from ppm to 100% by weight

Short analysis time (minutes)





Instrument calibration/drift correction.

a. Login a minimum of 3 Standard replicates.

b. Weigh a ~1.000 gram of a calibration/drift standard, enter the mass and standard identification into appropriate replicate fields.

c. Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open.

d. Place the calibration/drift standard into the open port at the top of the loading head.

e. Press the Analyze button on the instrument screen again, the loading head slide-block will close and the lower electrode will open.


国际**的非色散红外及热导检测器 ,无移动部件,不需要手动调整。

Instrument calibration/drift correction

f. Clean the upper and lower electrode manually, or, if applicable, remove the crucible and press the analyze button to clean with the automatic cleaner.

g. Firmly place a graphite crucible on the lower electrode tip.

h. Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically.

i. Repeat steps 3b through 3h a minimum of three times for each calibration/drift standard used.

. Calibrate/drift following the procedure outlined in the operator's instruction manual.


氧氮氢分析仪选哪家?给你推荐上海禹重科技!新疆LECO分析仪厂家供应

分析软件同时包括实时的环境参数的监测,提供完整的系统图表的模拟。山东元素分析仪咨询报价

氧氮氢分析仪的常见故障及解决方法

供电正常、通讯正常,点击确认键后分析仪不工作。这是没有水流,炉子温度太高或仪器通道电压不正常。如没有水流,炉子温度太高这些信息会显示在显示器画面上,但没有信息显示说明这两项正常。接下来检查仪器通道零位电压,如果比±3V高出1V以上,可能是因为气瓶空了,或者是空气进入到分析仪中。检查并更换化学试剂,如果有空气进入热导池里,热导池的电压就会<-6V,此时打开右面的门,堵住炉子气体进口,10s后,热导池电压值必然升高。经过逐一排查,**终确认碱石棉有问题,更换后仪器正常。 山东元素分析仪咨询报价

上海禹重实业有限公司位于友东路237号1幢104室。公司自成立以来,以质量为发展,让匠心弥散在每个细节,公司旗下光谱仪,色谱仪,质谱仪,样品前处理深受客户的喜爱。公司从事仪器仪表多年,有着创新的设计、强大的技术,还有一批**的专业化的队伍,确保为客户提供良好的产品及服务。禹重科技秉承“客户为尊、服务为荣、创意为先、技术为实”的经营理念,全力打造公司的重点竞争力。