氧氮氢分析仪的常见故障及解决方法
供电正常、通讯正常,点击确认键后分析仪不工作。这是没有水流,炉子温度太高或仪器通道电压不正常。如没有水流,炉子温度太高这些信息会显示在显示器画面上,但没有信息显示说明这两项正常。接下来检查仪器通道零位电压,如果比±3V高出1V以上,可能是因为气瓶空了,或者是空气进入到分析仪中。检查并更换化学试剂,如果有空气进入热导池里,热导池的电压就会<-6V,此时打开右面的门,堵住炉子气体进口,10s后,热导池电压值必然升高。经过逐一排查,**终确认碱石棉有问题,更换后仪器正常。 分析软件同时包括实时的环境参数的监测,提供完整的系统图表的模拟。元素分析仪厂家报价
Oxygen, Nitrogen, and Hydrogen in Refractory Metals
Summary
One of the most critical chemical specifications of titanium alloys is the hydrogen content. Too high of a
hydrogen content can cause hydrides to precipitate, which can lead to embrittlement and subsequent
cracking when the alloy is stressed. Hydrogen pickup typically occurs during downstream processing steps
such as heat treating, pickling, and cleaning.
The LECO ONH836 is a simultaneous oxygen, nitrogen, and hydrogen determinator that utilizes an electrode furnace, inert carrier gas, and both infrared and thermal conductivity detection to meet the analytical needs of the refractory metal industry.
This application note was written specifically for use with the LECO ONH836 series determinator.
氧氮氢分析仪是快速、准确分析氧、氮、氢元素的、高性能仪器,整机采用模块化、一体化设计,即主机由可靠的脉冲电极炉、高灵敏高精度的检测系统、布局合理和气密性好的气路系统、可靠耐用的电源系统、稳定的电路系统、高校节能的水循环散热系统等六个**的硬件模块单元组成。
氧氮氢分析仪的测量原理:
测量范围非常***。分析样品时,样品被称重后放到样品端口中,然后用载气进行冲洗,防止大气(含氧气和氮气)进入到炉子系统中。
石墨坩埚在脉冲炉中脱气,以尽量减少自带的污染。经过稳定阶段后样品落入坩埚中并熔融。样品中的氧与石墨坩埚中的碳反应生成一氧化碳。氮和氢以单质的形式释放。载气和样品气体通过粉尘过滤网后再进入到氧化铜催化炉中,将一氧化碳氧化成二氧化碳。
二氧化碳进入到红外池中用于测定氧含量,被测后的气体导入化学试剂管,这时二氧化碳和水被化学试剂除去,氮元素通过热导检测池含量被测定。测量氢时,载气换为氮气,样品气体通过舒茨试剂而不是氧化铜催化剂。测氧和氮时也可采用更廉价的氩气作为载气。
Procedure - Solid Samples
Prepare the instrument as outlined in the operator's instruction manual.
Determine the instrument blank.
Login a minimum of 3 Blank replicates.
Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open. Place a 502-344 Nickel Basket into the loading head.
Note: samples using automation should be placed in the appropriate autoloader position before starting the
analysis sequence. Once the sequence has started, the automatic analysis will start and end automatically.
ONH836氧氮氢分析仪
高频燃烧红外检测技分析无机材料中的碳、硫含量。
采用惰性气体保护熔融法,提供各种无机材料、钢铁、有色金属、硬质合金、难熔金属、陶瓷等材料中宽范围的氧氮氢同时分析。
该仪器配置MS Windows®的软件接口,专为触摸屏操作定制。长期的开发研究与客户反馈相结合,这个简单方便的软件使得用户可以快速完整访问仪器控制,分析设置,故障诊断,数据报告, **提高日常分析的效率并节省用户宝贵的时间。 可选的Cornerstone® Mobile软件功能,使得用户可以在手机,便携式电脑及台式电脑上便捷的访问仪器数据,分析曲线图形以及仪器的状态,实现对仪器的遥控。
减少了繁琐的额外数据处理。北京LECO分析仪厂家直销
为可**的Cornerstone 氧氮氧分析软件采用触摸屏界面使用户达到分析控制、15法设定、在线诊断、数据报告。元素分析仪厂家报价
Procedure – Powder/Chip Samples
Instrument calibration/drift correction.
Add approximately 0.05 g of *** Graphite Powder to a***Graphite Crucible.
Firmly place the crucible on the lower electrode tip or appropriate autoloader position.
Press the Analyze button on the instrument screen,the lower electrode will close and the analysis sequence will start and end automatically.
Repeat steps 3b through 3i a minimum of three times for each calibration/drift standard used.