首页 >  仪器仪表 >  安徽氧氮氢分析仪采购「上海禹重实业供应」

分析仪基本参数
  • 品牌
  • LECO
  • 型号
  • 齐全
分析仪企业商机

Instrument Highlights and Features


Features and Benefits

LED light ring illuminates the furnace area

Choice of either argon or helium carrier gas

Novel electrode design increases heat transfer efficiency for increased stability

Non-touch interface systems are also available

User-friendly brand software



High-Performance Detector Design

Thermostatic construction provides increased protection from ambient temperature fluctuations

Optimized emitter control and detection circuitryimproves the IR cell lifetime and long-term stability,

resulting in superior accuracy and precision

Infrared detection for the determination of oxygen, thermoconductivity detection for the determination

of nitrogen




可以旁路的OMI-2 气路强化过滤装置增强了检测器的稳定性,使得分析结果更加准确。安徽氧氮氢分析仪采购

氧氮氢分析仪的常见故障及解决方法:

氧和氮空白值超过20。这是由于气流小,不能将炉子中的空气驱赶出去。可调节气体流量,调节载气压力在0.2~0.4MPa。接通仪器载气,放一个石墨坩埚在下电极上,打开主电源开关,点击软件上的关炉按钮,关闭炉子并等待10s。调节流量调节器,直到流量计a显示为30L/h,打开炉子。调节调节器直到流量计b显示为50L/h,再次关闭炉子。如果以上设置不稳定,则增加流量至100L/h,反复调节直至仪器稳定。 浙江分析仪多少钱***触摸屏技术,用户友好界面软件进行校正、分析、诊断等。

ONH836系列:氧氮氢分析仪(熔融法)

力可ONH836系列仪器采用惰性熔融法,提供各种无机材料、钢铁、有色金属、硬质合金、陶瓷材料中宽范围的氧氮氢同时分析。

ONH836系列仪器通过***采纳客户的反馈意见和创新设计,力可***仪器采用触摸屏技术和在线软件平台,增强了仪器的可操作性、降低了分析成本、节省实验室桌面空间。


用户友好界面的Cornerstone™ 软件

力可**的Cornerstone 氧氮氧分析软件采用触摸屏界面使用户可完整的达到分析控制、方法设定、在线诊断、数据报告。Cornerstone  氧氮氧分析软件允许用户在单一分析界面中完成日常的所有工作,快速而且简单易用。力可创新的批次和重复样品数据分组功能极大简化了样品结果输出和结果标准偏差的自动计算,减少了繁琐的额外数据处理。


Procedure – Powder/Chip Samples

Instrument calibration/drift correction.

Add approximately 0.05 g of *** Graphite Powder to a***Graphite Crucible.

Firmly place the crucible on the lower electrode tip or appropriate autoloader position.

Press the Analyze button on the instrument screen,the lower electrode will close and the analysis sequence will start and end automatically.

Repeat steps 3b through 3i a minimum of three times for each calibration/drift standard used.

可选六轴可旋转触摸屏幕,增强了操作的工效和直观性。

氧氮氢分析仪的常见故障及解决方法

供电正常、通讯正常,点击确认键后分析仪不工作。这是没有水流,炉子温度太高或仪器通道电压不正常。如没有水流,炉子温度太高这些信息会显示在显示器画面上,但没有信息显示说明这两项正常。接下来检查仪器通道零位电压,如果比±3V高出1V以上,可能是因为气瓶空了,或者是空气进入到分析仪中。检查并更换化学试剂,如果有空气进入热导池里,热导池的电压就会<-6V,此时打开右面的门,堵住炉子气体进口,10s后,热导池电压值必然升高。经过逐一排查,**终确认碱石棉有问题,更换后仪器正常。 全自动分析系统(可选) 。浙江LECO氧氮氢分析仪

改进的红外检测池技术。安徽氧氮氢分析仪采购

Procedure - Solid Samples

Instrument calibration/drift correction.

Add approximately 0.05 g of *** Graphite Powder to a *** Graphite Crucible.

Firmly place the crucible on the lower electrode tip or appropriate autoloader position.

Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically.

Repeat steps 3b through 3j a minimum of three times for each calibration/drift standard used.

Calibrate/drift following the procedure outlined in the operator's instruction manual.


安徽氧氮氢分析仪采购

与分析仪相关的文章
与分析仪相关的产品
与分析仪相关的问题
与分析仪相关的搜索
信息来源于互联网 本站不为信息真实性负责