Software Features & Benefits
Cornerstone brand software is divided into four main sections—Analysis, Diagnostics, Settings, and Instrument—for simplified navigation and organization. Toolbars, sliders, and drop-down menus makeit easy to set parameters for calibration and data processing. The software also includes real-time ambient monitors, with fully animated system diagrams. Advanced diagnostic features include a thorough digital on-board manual, maintenance animations, photo illustrations, and screen captures that quickly provide the direction needed without having to refer to multiple manuals. also Cornerstone supports extended data archiving and flexible reporting capabilities. 工具条、滑动条和下拉式菜单简化了校正和数据处理的参数设定。新疆专业分析仪价格
Sample Preparation
Typically, titanium and zirconium samples are chemically etched to remove surface contamination when oxygen and nitrogen are determined. However, etching can introduce hydrogen into the sample.
ASTM method E 1409 "Determination of Oxygen and Nitrogen in Titanium and Titanium Alloys by the Inert
Gas Fusion Technique", as updated in 1996, permits either etching or abrading (filing) of the test specimen.
ASTM E 1447 "Determination of Hydrogen in Titanium and Titanium Alloys by the Inert Gas Fusion Thermal
Conductivity/Infrared Detection Method" permits surface preparation by abrading (if necessary to remove
contamination). Differences in sample preparation present somewhat of a dilemma regarding simultaneous
determination of O, N, and H in titanium. However, abrading samples with a file to remove surface
contamination will yield accurate O, N, and H results. The ONH836 utilizes a high-power electrode furnace to quickly and efficiently release the target gases from within the sample, which allows for a very rapid simultaneous determination of oxygen, nitrogen, and hydrogen.
广东美国分析仪采购高速全自动真空除尘装置直接将粉尘排至仪器外部捕集器中,消除环境污染。Instrument calibration/drift correction.
a. Login a minimum of 3 Standard replicates.
b. Weigh a ~1.000 gram of a calibration/drift standard, enter the mass and standard identification into appropriate replicate fields.
c. Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open.
d. Place the calibration/drift standard into the open port at the top of the loading head.
e. Press the Analyze button on the instrument screen again, the loading head slide-block will close and the lower electrode will open.
Oxygen, Nitrogen, and Hydrogen in Refractory Metals
Summary
One of the most critical chemical specifications of titanium alloys is the hydrogen content. Too high of a
hydrogen content can cause hydrides to precipitate, which can lead to embrittlement and subsequent
cracking when the alloy is stressed. Hydrogen pickup typically occurs during downstream processing steps
such as heat treating, pickling, and cleaning.
The LECO ONH836 is a simultaneous oxygen, nitrogen, and hydrogen determinator that utilizes an electrode furnace, inert carrier gas, and both infrared and thermal conductivity detection to meet the analytical needs of the refractory metal industry.
This application note was written specifically for use with the LECO ONH836 series determinator.
ONH836系列:氧氮氢分析仪(熔融法)
力可ONH836系列仪器采用惰性熔融法,提供各种无机材料、钢铁、有色金属、硬质合金、陶瓷材料中宽范围的氧氮氢同时分析。
ONH836系列仪器通过***采纳客户的反馈意见和创新设计,力可***仪器采用触摸屏技术和在线软件平台,增强了仪器的可操作性、降低了分析成本、节省实验室桌面空间。
用户友好界面的Cornerstone™ 软件
力可**的Cornerstone 氧氮氧分析软件采用触摸屏界面使用户可完整的达到分析控制、方法设定、在线诊断、数据报告。Cornerstone 氧氮氧分析软件允许用户在单一分析界面中完成日常的所有工作,快速而且简单易用。力可创新的批次和重复样品数据分组功能极大简化了样品结果输出和结果标准偏差的自动计算,减少了繁琐的额外数据处理。
高级诊断程序包括全数字化的在线操作手册、维护模拟演示、图片说明等提供快速指导。湖南LECO分析仪多少钱
提供各种无机材料、钢铁、有色金属、硬质合金、陶瓷材料中宽范围的氧氮氢同时分析。新疆专业分析仪价格
Procedure – Powder/Chip Samples
Instrument calibration/drift correction.
Calibrate/drift following the procedure outlined in the operator's instruction manual.
Login Sample with the appropriate number of reps.
Weigh approximately *** grams of appropriately prepared sample into a *** Nickel Capsule, enter mass and sample identification into appropriate rep fields. If applicable, place the nickel capsule in the
appropriate autoloader position.
Repeat steps 3c through 3i for sample analysis.