Procedure - Solid Samples

Prepare the instrument as outlined in the operator's instruction manual.


Determine the instrument blank.

Login a minimum of 3 Blank replicates.

Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open. Place a 502-344 Nickel Basket into the loading head.

Note: samples using automation should be placed in the appropriate autoloader position before starting the

analysis sequence. Once the sequence has started, the automatic analysis will start and end automatically.


长寿命的红外发射光源和无漂移的电路设计保证检测器的长期稳定性。北京元素分析仪咨询报价

Sample Preparation

Typically, titanium and zirconium samples are chemically etched to remove surface contamination when oxygen and nitrogen are determined. However, etching can introduce hydrogen into the sample.

ASTM method E 1409 "Determination of Oxygen and Nitrogen in Titanium and Titanium Alloys by the Inert

Gas Fusion Technique", as updated in 1996, permits either etching or abrading (filing) of the test specimen.

ASTM E 1447 "Determination of Hydrogen in Titanium and Titanium Alloys by the Inert Gas Fusion Thermal

Conductivity/Infrared Detection Method" permits surface preparation by abrading (if necessary to remove

contamination). Differences in sample preparation present somewhat of a dilemma regarding simultaneous

determination of O, N, and H in titanium. However, abrading samples with a file to remove surface

contamination will yield accurate O, N, and H results. The ONH836 utilizes a high-power electrode furnace to quickly and efficiently release the target gases from within the sample, which allows for a very rapid simultaneous determination of oxygen, nitrogen, and hydrogen.

湖南美国LECO分析仪价格改进的红外检测池技术。

氧氮氢分析仪的常见故障及解决方法:

氧和氮空白值超过20。这是由于气流小,不能将炉子中的空气驱赶出去。可调节气体流量,调节载气压力在0.2~0.4MPa。接通仪器载气,放一个石墨坩埚在下电极上,打开主电源开关,点击软件上的关炉按钮,关闭炉子并等待10s。调节流量调节器,直到流量计a显示为30L/h,打开炉子。调节调节器直到流量计b显示为50L/h,再次关闭炉子。如果以上设置不稳定,则增加流量至100L/h,反复调节直至仪器稳定。

美国LECO力可氧氮氢分析仪ONH836 氧氮氢联测仪系列软件部分

1、用户友好界面的Cornerstone软件

力可**的Cornerstone氧氮氢分析软件采用触摸屏界面使用户可完整的达到分析控制、方法设定、在线诊断、数据报告。Cornerstone氧氮氢分析软件允许用户在单一分析界面中完成日常的所有工作,快速而且简单易用。力可创新的批次和重复样品数据分组功能极大简化了样品结果输出和结果标准偏差的自动计算。减少了繁琐的额外数据处理。

2、软件特点和优势

Cornerstone氧氮氢分析软件分成四个主要部分:分析、诊断、设定、仪器,界面可快速导航和组织。工具条、滑动条和下拉式菜单简化了校正和数据处理的参数设定。分析软件同时包括实时的环境参数的检测,提供完整的系统图表模拟。高级诊断程序包括全数字化的在线操作手册、维护模拟演示、图片说明等提供快速指导。Cornerstone氧氮氢分析软件同时支持***的数据文档格式和可变的报告格式。


20位试样自动进样系统。

Procedure – Powder/Chip Samples

Prepare the instrument as outlined in the operator's instruction manual.


Determine the instrument blank.

Login a minimum of three Blank reps.

Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open. Insert a *** Nickel Capsule (leave capsule open) into the open port at top of loading head.

Note: samples using automation should be placed in the appropriate autoloader position before starting the analysis sequence. Once the sequence has started, the automatic analysis will start and end automatically.


试样加载过程中完全密闭,消除大气的中可能带来的杂质的污染。浙江专业分析仪价格

用户友好界面的Cornerstone™ 软件。北京元素分析仪咨询报价

Oxygen, Nitrogen, and Hydrogen in Refractory Metals

Summary

One of the most critical chemical specifications of titanium alloys is the hydrogen content. Too high of a

hydrogen content can cause hydrides to precipitate, which can lead to embrittlement and subsequent

cracking when the alloy is stressed. Hydrogen pickup typically occurs during downstream processing steps

such as heat treating, pickling, and cleaning.

The LECO ONH836 is a simultaneous oxygen, nitrogen, and hydrogen determinator that utilizes an electrode furnace, inert carrier gas, and both infrared and thermal conductivity detection to meet the analytical needs of the refractory metal industry. 

This application note was written specifically for use with the LECO ONH836 series determinator.