Procedure - Solid Samples
Prepare the instrument as outlined in the operator's instruction manual.
Determine the instrument blank.
Login a minimum of 3 Blank replicates.
Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open. Place a 502-344 Nickel Basket into the loading head.
Note: samples using automation should be placed in the appropriate autoloader position before starting the
analysis sequence. Once the sequence has started, the automatic analysis will start and end automatically.
Procedure – Powder/Chip Samples
Determine the instrument blank.
Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically.
Repeat steps 2b through 2g a minimum of three times.
Set the blank following the procedure outlined in the operator's instruction manual.
Instrument calibration/drift correction.
Login a minimum of three Standard reps.
Weigh approximately *** grams of a calibration/drift standard into a *** Nickel Capsule; enter the mass and standard identification into appropriate rep fields. If applicable, place the nickel capsule in the appropriate autoloader position.
广东氧氮氢分析仪厂家报价20位或100位全自动分析系统。
Procedure - Solid Samples
Instrument calibration/drift correction.
Add approximately 0.05 g of *** Graphite Powder to a *** Graphite Crucible.
Firmly place the crucible on the lower electrode tip or appropriate autoloader position.
Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically.
Repeat steps 3b through 3j a minimum of three times for each calibration/drift standard used.
Calibrate/drift following the procedure outlined in the operator's instruction manual.
氧氮氢分析仪的使用注意事项:
1 、氧氮氢分析仪在初次启用前,应该对连接点、焊点、阀门等进行检漏,以确保空气中的氧不会反渗进入管道及仪器内部,造成测量数值偏高。
2、再次使用仪器前,要进行管道系统净化,将漏入的空气除干净,同时确保连接取样管路时没有漏入空气。
3、样气中氧含量的变化会受管道材质及表面粗糙度影响,因此一般连接管路选用铜管或抛光过的不锈钢管,而不用塑料管、橡胶管等。
4、微量元素分析时,要避免各种管件、阀门、表头等死角对样气造成污染,因此必须尽可能的简化气路系统,连接件死角要小,以防止溶解氧逸出造成污染,使用水封、油封及腊封等设备,才能较好的确保数据的。
试样加载过程中完全密闭,消除大气的中可能带来的杂质的污染。
Oxygen, Nitrogen, and Hydrogen in Refractory Metals
Summary
One of the most critical chemical specifications of titanium alloys is the hydrogen content. Too high of a
hydrogen content can cause hydrides to precipitate, which can lead to embrittlement and subsequent
cracking when the alloy is stressed. Hydrogen pickup typically occurs during downstream processing steps
such as heat treating, pickling, and cleaning.
The LECO ONH836 is a simultaneous oxygen, nitrogen, and hydrogen determinator that utilizes an electrode furnace, inert carrier gas, and both infrared and thermal conductivity detection to meet the analytical needs of the refractory metal industry.
This application note was written specifically for use with the LECO ONH836 series determinator.
高速全自动真空除尘装置直接将粉尘排至仪器外部捕集器中,消除环境污染。美国LECO分析仪优惠价格
Procedure – Powder/Chip Samples
Instrument calibration/drift correction.
Calibrate/drift following the procedure outlined in the operator's instruction manual.
Login Sample with the appropriate number of reps.
Weigh approximately *** grams of appropriately prepared sample into a *** Nickel Capsule, enter mass and sample identification into appropriate rep fields. If applicable, place the nickel capsule in the
appropriate autoloader position.
Repeat steps 3c through 3i for sample analysis.