Procedure – Powder/Chip Samples
Instrument calibration/drift correction.
Calibrate/drift following the procedure outlined in the operator's instruction manual.
Login Sample with the appropriate number of reps.
Weigh approximately *** grams of appropriately prepared sample into a *** Nickel Capsule, enter mass and sample identification into appropriate rep fields. If applicable, place the nickel capsule in the
appropriate autoloader position.
Repeat steps 3c through 3i for sample analysis.
Software Features & Benefits
Cornerstone brand software is divided into four main sections—Analysis, Diagnostics, Settings, and Instrument—for simplified navigation and organization. Toolbars, sliders, and drop-down menus makeit easy to set parameters for calibration and data processing. The software also includes real-time ambient monitors, with fully animated system diagrams. Advanced diagnostic features include a thorough digital on-board manual, maintenance animations, photo illustrations, and screen captures that quickly provide the direction needed without having to refer to multiple manuals. also Cornerstone supports extended data archiving and flexible reporting capabilities. 浙江元素分析仪采购新型气路管道减少气路连接阀,使得漏气,断路故障更少。
Sample Preparation
Sampling and sample preparation of refractory metals such as titanium and zirconium is somewhat different from that of steel. Unlike steel samples, hydrogen is not as mobile in this group of materials; therefore, storage in liquid nitrogen or dry ice is not required. However, it is important to keep the sample cool when cutting or sectioning. Sample preparation for oxygen and nitrogen determination has beendifferent from that of hydrogen determination.
Instrument calibration/drift correction.
a. Login a minimum of 3 Standard replicates.
b. Weigh a ~1.000 gram of a calibration/drift standard, enter the mass and standard identification into appropriate replicate fields.
c. Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open.
d. Place the calibration/drift standard into the open port at the top of the loading head.
e. Press the Analyze button on the instrument screen again, the loading head slide-block will close and the lower electrode will open.
无需人工干预的全自动清扫系统。
Procedure – Powder/Chip Samples
Instrument calibration/drift correction.
Add approximately 0.05 g of *** Graphite Powder to a***Graphite Crucible.
Firmly place the crucible on the lower electrode tip or appropriate autoloader position.
Press the Analyze button on the instrument screen,the lower electrode will close and the analysis sequence will start and end automatically.
Repeat steps 3b through 3i a minimum of three times for each calibration/drift standard used.
***触摸屏技术,用户友好界面软件进行校正、分析、诊断等。专业元素分析仪厂家报价高性能的红外检测器配置***恒温绝热技术,消除了外界环境温度的干扰。浙江元素分析仪采购
氧氮氢分析仪
氧氮氢分析仪是一种用于材料科学领域的仪器,于2016年03月09日启用。
技术指标
测量范围: ON 低氧0.0-300ppm OH 低氧0.0-300ppm 高氧0.0-2% 高氧0.0-2% 低氮0.0―300ppm 低氢0.0-50ppm 高氮0.0-2% 高氢0.0-1000ppm 灵敏度: 氧0.01ppm;氮0.1ppm;氢0.01ppm。
主要功能
该仪器配置有16位微处理器和两个**的分别检测高氧和低氧的 红外检测池。氮和氢通过双重范围的热导检测池测量。分析过程中可自动实现从 低范围到高范围的切换。样品在高功率脉冲炉的石墨坩埚中加热可达3000℃以上 高温,脉冲炉采用循环水冷却。该仪器具有灵敏度高、性能好、测量范围宽和分 析结果准确可靠等优点。ONH-2000氧氮氢分析仪是为快速、准确测定铜、钢、铸 铁、合金、钛、镍、钼、锆、陶瓷和其它无机材料中氧、氮、氢的含量而专门设计 制造的。 浙江元素分析仪采购
产品保温效果可以得到充分保证。试验箱门与循环风机,提蓝传动等互锁,保护操作者的安全,一旦打开箱门,循...
【详情】电热恒温鼓风干燥箱如何养护?一)电热恒温鼓风干燥箱耗费的电流比较大。因为这个,它所用的电源线、闸刀开...
【详情】