Instrument calibration/drift correction
f. Clean the upper and lower electrode manually, or, if applicable, remove the crucible and press the analyze button to clean with the automatic cleaner.
g. Firmly place a graphite crucible on the lower electrode tip.
h. Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically.
i. Repeat steps 3b through 3h a minimum of three times for each calibration/drift standard used.
. Calibrate/drift following the procedure outlined in the operator's instruction manual.
User-Friendly Brand Software Cornerstone
LECO's exclusive brand software enables the user to have complete access to analysis Cornerstone control, method settings, diagnostics, reporting, and more in a highly organized and immersive environment. Designed through a collaboration of customer feedback and innovative engineering,Cornerstone allows the user to conduct all of their day-to-day operations within a single Analysis screen designed for speed and ease-of-use. Our innovative grouping of sample data into sets and replicates simplifies the data output and automatically calculates relevant statistics, alleviating the need for additional data processing.
北京美国分析仪厂家供应配置SmartLine® Remote远程诊断功能。
Oxygen, Nitrogen, and Hydrogen in Refractory Metals
Summary
One of the most critical chemical specifications of titanium alloys is the hydrogen content. Too high of a
hydrogen content can cause hydrides to precipitate, which can lead to embrittlement and subsequent
cracking when the alloy is stressed. Hydrogen pickup typically occurs during downstream processing steps
such as heat treating, pickling, and cleaning.
The LECO ONH836 is a simultaneous oxygen, nitrogen, and hydrogen determinator that utilizes an electrode furnace, inert carrier gas, and both infrared and thermal conductivity detection to meet the analytical needs of the refractory metal industry.
This application note was written specifically for use with the LECO ONH836 series determinator.
Procedure - Solid Samples
Instrument calibration/drift correction.
Place the calibration/drift standard in a *** Nickel Basket, and if applicable, place the sample into the appropriate autoloader position.
Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open.
Note: samples using automation should be placed in the appropriate autoloader position before starting the
analysis sequence. Once the sequence has started, the automatic analysis will start and end automatically.
company profile
Shanghai Uzong Industrial Co., Ltd. (hereinafter referred to as UZONGLAB®) is a high-tech company that provides integrated solutions for intelligent laboratories, mainly serving the Chinese market and countries along the “Belt and Road”. It has three business divisions and one The technology center, namely the instrument business department, consumables business department, project development department, and Zhongyu Lianhe Technology Center. In addition to the Shanghai headquarters, we have established branches in Beijing, Shenyang, Shijiazhuang, Taiyuan, Qingdao, Guangzhou, Chengdu and Urumqi. To provide customers with more professional, safer and more convenient products and services is our sole goal. 提供各种无机材料、钢铁、有色金属、硬质合金、陶瓷材料中宽范围的氧氮氢同时分析。美国分析仪厂家报价
20位试样自动进样系统。湖南美国LECO分析仪厂家供应
氧氮氢分析仪的使用注意事项:
1 、氧氮氢分析仪在初次启用前,应该对连接点、焊点、阀门等进行检漏,以确保空气中的氧不会反渗进入管道及仪器内部,造成测量数值偏高。
2、再次使用仪器前,要进行管道系统净化,将漏入的空气除干净,同时确保连接取样管路时没有漏入空气。
3、样气中氧含量的变化会受管道材质及表面粗糙度影响,因此一般连接管路选用铜管或抛光过的不锈钢管,而不用塑料管、橡胶管等。
4、微量元素分析时,要避免各种管件、阀门、表头等死角对样气造成污染,因此必须尽可能的简化气路系统,连接件死角要小,以防止溶解氧逸出造成污染,使用水封、油封及腊封等设备,才能较好的确保数据的。
湖南美国LECO分析仪厂家供应
液压万能试验机的油泵调试以及试验步骤:1、接好电源线,按“电源”按钮,指示灯亮。2、根据试样规格选择...
【详情】在这些领域中流量计主要是取到检查和测量的作用。能源计量。煤炭、原油、煤层气、石油气天然气等一次能源,...
【详情】现在,质量流量计在质量流量测试仪中处于主导地位,相信一直以这种理念发展下去的话,在这个行业中一直保持...
【详情】万能试验机人们往往将数据的稳定显示看的非常重要,而忽略了数据的真实性,将滤波器的截止频率设置的非常低...
【详情】万能试验机安装在清洁干燥而且温度均匀的房屋内同时应考虑到在机器上作较长的长梁弯曲试验的可能性以及使用...
【详情】电液伺服万能试验机集电液伺服自动控制、自动测量、数据采集、屏幕显示、试验结果处理为一体,以油缸下置式...
【详情】