Sample Preparation
Sampling and sample preparation of refractory metals such as titanium and zirconium is somewhat different from that of steel. Unlike steel samples, hydrogen is not as mobile in this group of materials; therefore, storage in liquid nitrogen or dry ice is not required. However, it is important to keep the sample cool when cutting or sectioning. Sample preparation for oxygen and nitrogen determination has beendifferent from that of hydrogen determination.
20位试样自动进样系统。山西元素分析仪厂家直销
Procedure – Powder/Chip Samples
Instrument calibration/drift correction.
Calibrate/drift following the procedure outlined in the operator's instruction manual.
Login Sample with the appropriate number of reps.
Weigh approximately *** grams of appropriately prepared sample into a *** Nickel Capsule, enter mass and sample identification into appropriate rep fields. If applicable, place the nickel capsule in the
appropriate autoloader position.
Repeat steps 3c through 3i for sample analysis.
Procedure - Solid Samples
Press the Analyze button on the instrument screen again, the loading head slide-block will close and the lower electrode will open.
Clean the upper and lower electrode manually, or, if applicable, remove the crucible and press the analyze button to clean with the automatic cleaner
Add approximately 0.05 g of *** Graphite Powder to a *** Graphite Crucible.
Firmly place the crucible on the lower electrode tip or appropriate autoloader position.
Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically
Instrument calibration/drift correction
f. Clean the upper and lower electrode manually, or, if applicable, remove the crucible and press the analyze button to clean with the automatic cleaner.
g. Firmly place a graphite crucible on the lower electrode tip.
h. Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically.
i. Repeat steps 3b through 3h a minimum of three times for each calibration/drift standard used.
. Calibrate/drift following the procedure outlined in the operator's instruction manual.
Procedure - Solid Samples
Instrument calibration/drift correction.
Add approximately 0.05 g of *** Graphite Powder to a *** Graphite Crucible.
Firmly place the crucible on the lower electrode tip or appropriate autoloader position.
Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically.
Repeat steps 3b through 3j a minimum of three times for each calibration/drift standard used.
Calibrate/drift following the procedure outlined in the operator's instruction manual.