首页 >  仪器仪表 >  山西元素分析仪厂家直销「上海禹重实业供应」

分析仪基本参数
  • 品牌
  • LECO
  • 型号
  • 齐全
分析仪企业商机

Sample Preparation


Sampling and sample preparation of refractory metals such as titanium and zirconium is somewhat different from that of steel. Unlike steel samples, hydrogen is not as mobile in this group of materials; therefore, storage in liquid nitrogen or dry ice is not required. However, it is important to keep the sample cool when cutting or sectioning. Sample preparation for oxygen and nitrogen determination has beendifferent from that of hydrogen determination.



20位试样自动进样系统。山西元素分析仪厂家直销

Procedure – Powder/Chip Samples

Instrument calibration/drift correction.

Calibrate/drift following the procedure outlined in the operator's instruction manual.


Analyze Samples.

Login Sample with the appropriate number of reps.

Weigh approximately *** grams of appropriately prepared sample into a *** Nickel Capsule, enter mass and sample identification into appropriate rep fields. If applicable, place the nickel capsule in the

appropriate autoloader position.

Repeat steps 3c through 3i for sample analysis.


元素分析仪厂家报价上海禹重给您提供各种氧氮氢分析仪,欢迎来电咨询!

Procedure - Solid Samples

Press the Analyze button on the instrument screen again, the loading head slide-block will close and the lower electrode will open.


Clean the upper and lower electrode manually, or, if applicable, remove the crucible and press the analyze button to clean with the automatic cleaner


Add approximately 0.05 g of *** Graphite Powder to a *** Graphite Crucible.


Firmly place the crucible on the lower electrode tip or appropriate autoloader position.


Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically



Instrument calibration/drift correction

f. Clean the upper and lower electrode manually, or, if applicable, remove the crucible and press the analyze button to clean with the automatic cleaner.

g. Firmly place a graphite crucible on the lower electrode tip.

h. Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically.

i. Repeat steps 3b through 3h a minimum of three times for each calibration/drift standard used.

. Calibrate/drift following the procedure outlined in the operator's instruction manual.


试样加载过程中完全密闭,消除大气的中可能带来的杂质的污染。

Procedure - Solid Samples

Instrument calibration/drift correction.

Add approximately 0.05 g of *** Graphite Powder to a *** Graphite Crucible.

Firmly place the crucible on the lower electrode tip or appropriate autoloader position.

Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically.

Repeat steps 3b through 3j a minimum of three times for each calibration/drift standard used.

Calibrate/drift following the procedure outlined in the operator's instruction manual.


新型气路管道减少气路连接阀,使得漏气,断路故障更少。上海专业元素分析仪厂家供应

**的动态流量补偿(DFC)保证热导池气体流量、压力恒定。山西元素分析仪厂家直销

ONH836系列:氧氮氢分析仪(熔融法)

力可ONH836系列仪器采用惰性熔融法,提供各种无机材料、钢铁、有色金属、硬质合金、陶瓷材料中宽范围的氧氮氢同时分析。

ONH836系列仪器通过***采纳客户的反馈意见和创新设计,力可***仪器采用触摸屏技术和在线软件平台,增强了仪器的可操作性、降低了分析成本、节省实验室桌面空间。


用户友好界面的Cornerstone™ 软件

力可**的Cornerstone 氧氮氧分析软件采用触摸屏界面使用户可完整的达到分析控制、方法设定、在线诊断、数据报告。Cornerstone  氧氮氧分析软件允许用户在单一分析界面中完成日常的所有工作,快速而且简单易用。力可创新的批次和重复样品数据分组功能极大简化了样品结果输出和结果标准偏差的自动计算,减少了繁琐的额外数据处理。


山西元素分析仪厂家直销

与分析仪相关的文章
与分析仪相关的产品
与分析仪相关的问题
与分析仪相关的搜索
信息来源于互联网 本站不为信息真实性负责