Procedure - Solid Samples

Instrument calibration/drift correction.

Place the calibration/drift standard in a *** Nickel Basket, and if applicable, place the sample into the appropriate autoloader position.

Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open.

Note: samples using automation should be placed in the appropriate autoloader position before starting the

analysis sequence. Once the sequence has started, the automatic analysis will start and end automatically.



装有自动旁路阀的可快速拆装氧水捕集装置。安徽专业元素分析仪厂家供应

Instrument Highlights and Features


Features and Benefits

LED light ring illuminates the furnace area

Choice of either argon or helium carrier gas

Novel electrode design increases heat transfer efficiency for increased stability

Non-touch interface systems are also available

User-friendly brand software



High-Performance Detector Design

Thermostatic construction provides increased protection from ambient temperature fluctuations

Optimized emitter control and detection circuitryimproves the IR cell lifetime and long-term stability,

resulting in superior accuracy and precision

Infrared detection for the determination of oxygen, thermoconductivity detection for the determination

of nitrogen




安徽专业元素分析仪厂家供应提供各种无机材料、钢铁、有色金属、硬质合金陶瓷材料中宽范围的氧氮氢同时测定。

力可OHN736&836系列仪器采用惰性熔融法,提供各种无机材料、钢铁、有色金属、硬质合金陶瓷材料中宽范围的氧氮氢同时测定。ONH836仪器通过***采纳客户的反馈意见和创新设计,力可***仪器采用触摸屏技术和在线软件平台,增强了仪器的可操作性,降低了分析成本,节省实验室桌面空间。


用户友好界面的Cornerstone™ 软件

力可**的Cornerstone碳硫分析软件采用触摸屏界面使用户可完整的达到分析控制、方法设定、在线诊断、数据报告。Cornerstone碳硫分析软件允许用户在单一分析界面中完成日常的所有工作,快速而且简单易用。力可创新的批次和重复样品数据分组功能极大简化了样品结果输出和结果标准偏差的自动计算。减少了繁琐的额外数据处理。


Sample Preparation

Typically, titanium and zirconium samples are chemically etched to remove surface contamination when oxygen and nitrogen are determined. However, etching can introduce hydrogen into the sample.

ASTM method E 1409 "Determination of Oxygen and Nitrogen in Titanium and Titanium Alloys by the Inert

Gas Fusion Technique", as updated in 1996, permits either etching or abrading (filing) of the test specimen.

ASTM E 1447 "Determination of Hydrogen in Titanium and Titanium Alloys by the Inert Gas Fusion Thermal

Conductivity/Infrared Detection Method" permits surface preparation by abrading (if necessary to remove

contamination). Differences in sample preparation present somewhat of a dilemma regarding simultaneous

determination of O, N, and H in titanium. However, abrading samples with a file to remove surface

contamination will yield accurate O, N, and H results. The ONH836 utilizes a high-power electrode furnace to quickly and efficiently release the target gases from within the sample, which allows for a very rapid simultaneous determination of oxygen, nitrogen, and hydrogen.

ONH836仪器通过***采纳客户的反馈意见和创新设计。

Procedure – Powder/Chip Samples

Determine the instrument blank.

Press the Analyze button on the instrument screen again, the loading head slide-block will close and the lower electrode will open.

Clean the upper and lower electrode either manually or remove the crucible and press the analyze button to clean with an automatic cleaner if applicable.

Add approximately 0.05 g of *** Graphite Powder to a 782-720 Graphite Crucible.

Firmly place the crucible on the lower electrode tip or appropriate autoloader position.


氧氮氢分析仪选哪家?给你推荐上海禹重科技!广东氧氮氢分析仪供应商

***电极炉头设计,提高脉冲炉热交换效率。安徽专业元素分析仪厂家供应

Method Selection


Two methods are described in this application note; either method can be used to analyze iron, steel, nickel-, and cobalt-base alloys. The Precision Method is recommended for general use and will provide the best precision and accuracy throughout the typical O, N, and H concentrations found in this group of metals; approximate cycle time is 3.5 minutes. The Fast Track method can be used where speed of analysis is a critical component; for example, when molten metal is being sampled and results are required in the shortest possible time. This method will produce suitable results for most samples; approximate cycle time is 2.25 minutes. As noted above, sampling and sample preparation are key elements to accurate O, N, and Hdetermination as well. It is up to the user to determine which method best meets their needs.