Determine the instrument blank.
a. Login a minimum of 3 Blank replicates.
b. Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open.
c. Press the Analyze button on the instrument screen again, the loading head slide-block will close and the lower electrode will open.
d. Clean the upper and lower electrode manually, or, if applicable, remove thecrucible and press the analyze button to clean with the automatic cleaner.
e. Firmly place a graphite crucible on the lower electrode tip.
f. Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and endautomatically.
g. Repeat steps 2b through 2f a minimum of three times.
h. Set the blank following the procedure outlined in the operator's instruction manual.
试样加载过程中完全密闭,消除大气的中可能带来的杂质的污染。山东氧氮氢分析仪哪家好
company profile
Shanghai Uzong Industrial Co., Ltd. (hereinafter referred to as UZONGLAB®) is a high-tech company that provides integrated solutions for intelligent laboratories, mainly serving the Chinese market and countries along the “Belt and Road”. It has three business divisions and one The technology center, namely the instrument business department, consumables business department, project development department, and Zhongyu Lianhe Technology Center. In addition to the Shanghai headquarters, we have established branches in Beijing, Shenyang, Shijiazhuang, Taiyuan, Qingdao, Guangzhou, Chengdu and Urumqi. To provide customers with more professional, safer and more convenient products and services is our sole goal. 浙江分析仪供应商氧氮氢分析仪选哪家?给您推荐上海禹重!
氧氮氢分析仪的操作过程
氧氮氢分析仪可根据的用户的要求定制。该分析仪配备两个不同灵敏度的探测器,在更多的应用中保证了准确度。
操作简单而且安全。在样品称重以后,样品的重量数据通过电子界面传导至相连的电脑中,也可以在软件界面中手动输入样品的重量。
样品被放置在水平放置的炉子中的低温区域。分析开始以后,炉子开始向上转动,然后样品逐渐落入加热区。通过添加载体气体氮气,氢气被扩散出来并且随载体被传导至一个高灵敏度的热传导测量池。
通常一次测量的时间在3到15分钟之间。在分析的过程中设备可以显示测试信号和设备的参数,从信号的评估到结果在界面上的显示全部自动完成;而测量数据则被传导至LIMS(实验室信息管理系统)中。无需过多的保养和维护,而除尘器和净化管的维护工作也方便。
Procedure - Solid Samples
Determine the instrument blank.
Repeat steps 2b through 2g a minimum of three times.
Set the blank following the procedure outlined in the operator's instruction manual.
Instrument calibration/drift correction.
Login a minimum of 3 Standard replicates.
Weigh approximately *** grams of a calibration/drift standard, enter the mass and standard identification into appropriate replicate fields.
Note: LECO Reference Materia***o not require preparation.See preparation statement on the reference material certificate.
装有自动旁路阀的可快速拆装氧水捕集装置。
Method Selection
Two methods are described in this application note; either method can be used to analyze iron, steel, nickel-, and cobalt-base alloys. The Precision Method is recommended for general use and will provide the best precision and accuracy throughout the typical O, N, and H concentrations found in this group of metals; approximate cycle time is 3.5 minutes. The Fast Track method can be used where speed of analysis is a critical component; for example, when molten metal is being sampled and results are required in the shortest possible time. This method will produce suitable results for most samples; approximate cycle time is 2.25 minutes. As noted above, sampling and sample preparation are key elements to accurate O, N, and Hdetermination as well. It is up to the user to determine which method best meets their needs.
提供各种无机材料、钢铁、有色金属、硬质合金、陶瓷材料中宽范围的氧氮氢同时分析。美国分析仪哪家好
配置SmartLine® Remote远程诊断功能。山东氧氮氢分析仪哪家好
Instrument Highlights and Features
Optional Features
Optional boom-mounted touch-screen interface package (M) promotes an ergonomic workspace
Optional Performance Package (P) adds a three-stage carrier gas purification system, and a dual-loop gas dose module
Optional Dual Cooling Package (D) adds an integratedliquid-to-air radiator with dual cooling fans. This DC
option gives the user the choice of liquid-to-liquid cooling, liquid-to-air cooling, or a combination of both
Optional integrated autocleaner (H) automatically cleans the upper electrode and lower electrode tip
Optional 20-sample shuttle loader available
山东氧氮氢分析仪哪家好
上海新漫致力于成为国际的核辐射测量、安全检查设备、系统供应商。新漫是一家集自主研发、生产、销售和服务...
【详情】α粒子在物质中的射程极短,因此对α粒子的探测必须使用薄窗式探测器。离子注入硅探测器(PIPS)为薄窗...
【详情】新漫项目根据流程执行。设有《项目任务书》。其中涵盖项目中对应阶段的技术指标要求,产品适用性,满足用户...
【详情】新漫项目根据流程执行。设有《项目任务书》。其中涵盖项目中对应阶段的技术指标要求,产品适用性,满足用户...
【详情】上海新漫传感技术研究发展有限公司(简称“新漫传感”)是一家服务核电、环境和公共安全领域的****,致...
【详情】PA系列只需点击“文件”→“优先项”→“保存路径”→“样品测试文件”,就可以设置测试数据的默认保存位...
【详情】新漫产品设有质量部门,有严格的质量标准,新漫质量体系符合ISC9001:2015标准。实施ISO90...
【详情】新漫实行的环境管理体系是ISO14001标准,是由ISO/TC207制订的环境管理体系标准之一。IS...
【详情】新漫项目根据流程执行。设有《项目任务书》。其中涵盖项目中对应阶段的技术指标要求,产品适用性,满足用户...
【详情】新漫传感一贯坚持精诚合作、和谐创新的经营之路,共谋发展企业的辉煌事业,起步于科研、院校的实验仪器、教...
【详情】中国产业调研网发布的2019-2025年中国安检设备市场深度调查分析及发展趋势研究报告认为,在安防行...
【详情】