氧氮氢分析仪的常见故障及解决方法
供电正常、通讯正常,点击确认键后分析仪不工作。这是没有水流,炉子温度太高或仪器通道电压不正常。如没有水流,炉子温度太高这些信息会显示在显示器画面上,但没有信息显示说明这两项正常。接下来检查仪器通道零位电压,如果比±3V高出1V以上,可能是因为气瓶空了,或者是空气进入到分析仪中。检查并更换化学试剂,如果有空气进入热导池里,热导池的电压就会<-6V,此时打开右面的门,堵住炉子气体进口,10s后,热导池电压值必然升高。经过逐一排查,**终确认碱石棉有问题,更换后仪器正常。 提供各种无机材料、钢铁、有色金属、硬质合金、陶瓷材料中宽范围的氧氮氢同时分析。浙江美国分析仪采购
Procedure – Powder/Chip Samples
Instrument calibration/drift correction.
Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open.
Place the Nickel Capsule containing the sample into the open port at the top of the loading head.
Press the Analyze button on the instrument screen again, the loading head slide-block will close and the lower electrode will open.
Clean the upper and lower electrode either manually or remove the crucible and press the analyze button to clean with an automatic cleaner if applicable.
Procedure - Solid Samples
Determine the instrument blank.
Repeat steps 2b through 2g a minimum of three times.
Set the blank following the procedure outlined in the operator's instruction manual.
Instrument calibration/drift correction.
Login a minimum of 3 Standard replicates.
Weigh approximately *** grams of a calibration/drift standard, enter the mass and standard identification into appropriate replicate fields.
Note: LECO Reference Materia***o not require preparation.See preparation statement on the reference material certificate.
氧氮氢分析仪的原理:分析样品在惰性气流存在下于石墨坩埚中加热熔融,其中脉冲炉温度可自由设定,并通过一个非接触式的光学温度传感器进行实时监控,用于实现样品的完全分解,反应所生成的CO H2和N2被带入到具有高稳定性和灵敏度的检测系统进行检测。对于CO的检测采用的是非色散性红外检测器,对于N2和H2的检测则采用热导检测器。
氧氮氢分析仪的特点:快速准确分析,自动化操作;精密气体校准;CO红外检测器,高分辨率和重现性的N、H热导池检测器;同一台仪器实现熔融提取和热提取;热提取分析时间可调;可自由选择温度*高达2500 °C;光学温度计温度测定和控制;所有检测器自动调零/自动水平控制(ALC);数据评估的自动优化选择范围;炉自动清洁和坩埚自动进样(可选);外设电脑进行操作和控制;样品信息保存; 所有样品数据可追踪查询;通过FTP或本地局域网进行数据传输; ON,H分析轻松切换。
试样加载过程中完全密闭,消除大气的中可能带来的杂质的污染。
Instrument calibration/drift correction
f. Clean the upper and lower electrode manually, or, if applicable, remove the crucible and press the analyze button to clean with the automatic cleaner.
g. Firmly place a graphite crucible on the lower electrode tip.
h. Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically.
i. Repeat steps 3b through 3h a minimum of three times for each calibration/drift standard used.
. Calibrate/drift following the procedure outlined in the operator's instruction manual.
可选六轴可旋转触摸屏幕,增强了操作的工效和直观性。浙江美国分析仪采购
LECO—Your source for total analytical solutions
CS CS 844 and 744 Series: Carbon/Sulfur by Combustion
Carbon and sulfur determination by combustion infrared detection
Quick, accurate, and affordable determination for both production control and research
Calibration, analysis, evaluation, and diagnostic functions accessible via user-friendly brand software
GDS GDS 500A and 850A Atomic Emission Spectrometers
Spectral range from 120 to 800 nm
Continuous profile of concentration/thickness
Large, dynamic range with concentrations from ppm to 100% by weight
Short analysis time (minutes)
浙江美国分析仪采购
中国产业调研网发布的2019-2025年中国安检设备市场深度调查分析及发展趋势研究报告认为,在安防行...
【详情】新漫售后服务部能为用户提供周到及时的保修和技术支持服务。如果因为产品设计或者生产装配质量的原因造成的...
【详情】技术特性探测器:2英寸碘化钠晶体(可选配1英寸氟溴化镧晶体或1.5英寸氟氯化镧晶体)内置GM管探测器...
【详情】上海新漫传感研究发展有限公司自2003年成立以来,一直专注于推动核辐射测量和快速安检相关技...
【详情】ISO9001标准涉及了技术法规、标准和合格评定程序。世贸/技术壁垒协定(WTO/TBT)也是WTO...
【详情】上海新漫致力于成为国际**的核辐射测量、安全检查设备、系统供应商。新漫是一家集自主研发、生产、销售和...
【详情】新漫售后服务部能为用户提供周到及时的保修和技术支持服务。如果因为产品设计或者生产装配质量的原因造成的...
【详情】上海新漫秉承“以科技为动力、以质量求发展、以信誉拓市场,为了更安全的明天”的经营理念并以开拓创新的思...
【详情】谱分析中效率刻度是定量分析的重要前提。效率刻度后可对所测样品进行定性分析。用户可以填写效率刻度相关信...
【详情】新漫Alpha谱仪系列与其他品牌的不同在于采用了离子注入型硅探测器(PIPS);探测室、前放、主放、...
【详情】