Instrument Highlights and Features


Optional Features

Optional boom-mounted touch-screen interface package (M) promotes an ergonomic workspace

Optional Performance Package (P) adds a three-stage carrier gas purification system, and a dual-loop gas dose module

Optional Dual Cooling Package (D) adds an integratedliquid-to-air radiator with dual cooling fans. This DC

option gives the user the choice of liquid-to-liquid cooling, liquid-to-air cooling, or a combination of both

Optional integrated autocleaner (H) automatically cleans the upper electrode and lower electrode tip

Optional 20-sample shuttle loader available



20位或100位全自动分析系统。进口分析仪

Procedure – Powder/Chip Samples

Determine the instrument blank.


Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically.


Repeat steps 2b through 2g a minimum of three times.


Set the blank following the procedure outlined in the operator's instruction manual.


Instrument calibration/drift correction.

Login a minimum of three Standard reps.


Weigh approximately *** grams of a calibration/drift standard into a *** Nickel Capsule; enter the mass and standard identification into appropriate rep fields. If applicable, place the nickel capsule in the appropriate autoloader position.


Note: LECO Reference Materia***o not require preparation. See preparation statement on the reference material certificate. Solid Standards may be used to calibrate when chip or powder standards are not available.




广东LECO分析仪厂家直销Cornerstone氧氮氢分析软件同时支持***的数据文档格式和可变的报告格式。

力可ONH836氧氮氢分析仪

特点和优势

国际**的非色散红外及热导检测器 ,无移动部件,不需要手动调整

采用三个**的红外器检测氧含量,同时检测CO和CO2,然后在转化为CO2集中检测,原理上是**完善的,所以不仅可以保证中间量程的测量精度,在**含量和高含量的检测范围内同样可以满足精度要求

高性能的红外检测器配置***恒温绝热技术,消除了外界环境温度的干扰

长寿命的红外发射光源和无漂移的电路设计保证检测器的长期稳定性

力可**的动态流量补偿控制DFC系统,保证了流量系统的完善,可以精确控制整个气路中的流量,保持流量恒定**改善了材料中氮分析的准确度

可以旁路的OMI-2 气路强化过滤装置增强了检测器的稳定性,使得分析结果更加准确

可选自动功能

20位试样自动进样系统

通用下电极坩埚支座,容纳各种坩埚使用

无需人工干预的全自动清扫系统



Procedure - Solid Samples

Prepare the instrument as outlined in the operator's instruction manual.


Determine the instrument blank.

Login a minimum of 3 Blank replicates.

Press the Analyze button on the instrument screen. After a short delay, the loading head slide-block will open. Place a 502-344 Nickel Basket into the loading head.

Note: samples using automation should be placed in the appropriate autoloader position before starting the

analysis sequence. Once the sequence has started, the automatic analysis will start and end automatically.


强力内循环水冷装置。

Method Selection


Two methods are described in this application note; either method can be used to analyze iron, steel, nickel-, and cobalt-base alloys. The Precision Method is recommended for general use and will provide the best precision and accuracy throughout the typical O, N, and H concentrations found in this group of metals; approximate cycle time is 3.5 minutes. The Fast Track method can be used where speed of analysis is a critical component; for example, when molten metal is being sampled and results are required in the shortest possible time. This method will produce suitable results for most samples; approximate cycle time is 2.25 minutes. As noted above, sampling and sample preparation are key elements to accurate O, N, and Hdetermination as well. It is up to the user to determine which method best meets their needs.








**的动态流量补偿(DFC)保证热导池气体流量、压力恒定。上海LECO分析仪采购

工具条、滑动条和下拉式菜单简化了校正和数据处理的参数设定。进口分析仪

氧氮氢分析仪的常见故障及解决方法

仪器启动时显示。没有水流。系统分析电流切断,分析停止。这是水流探测器不正常,水泵不工作,管道堵塞。将仪器的右面板取下,观察水流探测器,用手挤压补水塑料水瓶,发现水流正常,显示正常,证明水流探测器正常,管道畅通。启动循环水泵,但分析仪显示没有水流,此时可判定水泵不正常。打开水泵转子密封口,启动泵发现电机正常运转,此时关闭进水,拆下水泵,发现叶轮脱落。经了解,判定是由于外部冷却水停水,仪器内循环水温过高(水温应≤70℃),致使叶轮(叶轮材料PVC)热胀并脱轴。用粘合剂粘合叶轮后再粘于叶轮轴上,待粘合剂凝固后试车,仪器运行正常。 进口分析仪